The Delta-T DL6 is a battery-powered data logger in an IP67 case, built for soil moisture monitoring. It logs one Profile Probe or up to six soil moisture sensors, plus a 2K thermistor and a rain gauge, has a relay output and stores up to 16,000 readings at intervals of 1 s to 24 h.
The DL6 is a nine-channel data logger from Delta-T Devices dedicated to soil moisture monitoring. It offers six differential analogue channels for soil moisture sensors, one resistance channel for a 2K thermistor, one event counter channel for a rain gauge and one versatile relay channel, and it stores up to 16,000 readings.
It works with one Profile Probe (PR2/4, PR2/6, PR1/4 or PR1/6), or with up to six SM series sensors (SM150, SM150T, SM200, SM300) or ThetaProbes (ML2, ML3), together with one temperature sensor and one rain gauge (RG1 or RG2); two additional soil moisture sensors can be used alongside a PR2/4 or PR1/4. The DL6 handles a PR2/6 (100 cm depth) or a PR2/4 (40 cm depth), which plug in with the IP68 PR2 to DL6 cable (PRC/M12-05) with no wiring required. EQ3 sensors are also supported.
Designed for field use in research and irrigation monitoring, the DL6 runs on six internal AA alkaline batteries for up to one year, depending on logging frequency. Its status is checked by tapping the case and reading the status LED, and data is collected through the external RS232 connector with a laptop, or remotely via modem. Accelerated logging, triggered by the rain gauge or an event, helps track wetting fronts, and up to 10 DL6 loggers can share communications over an M8 cabling network.
| Manufacturer | Delta-T Devices Ltd |
|---|---|
| Model | DL6 |
| Channels | 9: 6 differential analogue, 1 resistance (2K thermistor), 1 event counter, 1 relay |
| Analogue channels (soil moisture) | 6; 0.0 – 0.1 m³.m⁻³: ±0.001 m³.m⁻³; 0.1 – 0.6 m³.m⁻³: ±1% of reading; applicable to ML3, SM300, PR2 (with up to 100 m cable) and PR1 (up to 25 m cable); mineral and organic soil calibrations |
| Analogue channels (DC voltage) | 6; -10 to 1150 mV; accuracy ±(0.3 mV + 0.2% of reading) + common mode error 0.25 mV/V; common mode range -0.2 to 3 V; logger reads either volumetric water content or voltage (not both) |
| Analogue input signal | Nominally 0 to 1 V; full signal range -0.01 to +1.15 V; permitted input voltage range -0.2 to +3 V relative to logger ground |
| Temperature channel | 1; sensor -20 to +60 °C; accuracy ±0.4 °C; ST1, ST2 or other 2K thermistor |
| Accuracy reference conditions | Logger at -10 to +50 °C |
| Event counter channel | 1; <50 Hz, contact closure or pulse logic inputs; max 1 mm/second with RG2 rain gauge (0.2 mm/tip) |
| Sensor excitation | 6; 5 – 9 V at 120 mA from internal battery voltage supply (unregulated); warm-up time in multiples of 1 s |
| Alarm output | 1; 1 A, 25 V DC/AC; isolated relay rated 2 A, protected by 1 A fuse |
| Logging frequency | 1 s to 24 h, user selectable |
| Accelerated logging | 1 s to 24 h, rain gauge or event triggered |
| Reading capacity | 16k readings; auto-wrap or stop logging, user selectable |
| PC connection | 1 × RS232 (USB to RS232 converters available); modem option |
| Software | DeltaLINK-PC version 3.6 or later; Windows XP, Vista, 7, 8, 10 or later |
| Status indicator | LED, activated by tapping the logger |
| Power | 6 × AA alkaline cells; lithium batteries for extended life and full temperature range |
| Battery lifetime | Up to 1 year (depends on logging frequency); lithium typically 3 to 5 times longer |
| Enclosure rating | IP67 |
| Operating temperature | -20 to +60 °C; full accuracy -10 to +50 °C |
| Compliance | CE, EN61326 (1997) |
| Size | 180 × 160 × 70 mm |