The Delta-T Devices GP2 is a weatherproof (IP65), research-grade data logger and controller with 12 differential analog channels, 4 digital inputs, SDI-12 for up to 62 sensors, relay control outputs, a Script Editor, a program Simulator and full Penman-Monteith ETo calculation.
The Delta-T Devices GP2 is a powerful, weatherproof, research-grade data logger with unique features for recording and controlling field experiments. It provides 12 differential (or 24 single-ended) analog channels with microvolt sensitivity, 4 digital inputs, a dedicated Delta-T WET Sensor channel and unlimited virtual channels. Since March 2016 every GP2 is SDI-12 enabled as standard, adding up to 62 SDI-12 sensors while all analog and digital channels remain available; earlier units are upgraded free of charge with the latest DeltaLINK software.
Set-up is point-and-click in the supplied DeltaLINK PC software. Each analog channel has its own input type and recording parameters, with 4 input ranges, adaptive auto-ranging and configurable sensor excitation, so nearly all analog sensors can be used and custom sensor types can be added to the library. The Script Editor builds control sequences without any programming language – irrigation and PID control, disease prediction, degree days, dew point or wind chill – and control parameters can be adjusted without interrupting logging. The Simulator tests a program before deployment, simulating years of logging in a few minutes. With DeltaLINK 3.7 and later, the GP2 calculates the full ASCE/FAO-56 Penman-Monteith reference evapotranspiration (ETo) from relative humidity, wind speed, solar radiation and air temperature, as hourly and daily values.
For the field, 4 MB of FLASH memory stores 2.5 million readings (typical); data are collected locally via USB/RS232 or remotely through cellular modem options, and can be viewed and shared on the DeltaLINK-Cloud platform. The logger runs on 6 internal AA alkaline batteries, with the optional GP2-PSU mains supply, and up to 7 GP2s can share power and communications over M12 network cabling; GP1 loggers can join such a network via an adapter cable. Typical fields include agronomy, meteorology, hydrology, eco-physiology and soil science. Delta-T’s compatibility list gives the maximum sensors per GP2 as: ML3, SM150T and EQ3 – 6 with temperature or 12 without; analogue PR2 – 2 PR2/6 or 3 PR2/4; PR2 SDI-12 – 50 PR2/6 or 62 PR2/4; WET150 – 62; temperature sensors and tensiometers – 12 each.
| Manufacturer | Delta-T Devices Ltd (UK) |
|---|---|
| Model | GP2 |
| Analog inputs | 12 differential inputs, configurable as a combination of: differential voltage (12 max), single-ended voltage (common ground, 24 max), 2-wire resistance (24 max), 3-wire resistance (12 max), bridge & potentiometric sensors (12 max), temperature sensors (12 max; 2-wire thermistors 24 max) |
| Temperature sensors | Thermistors (types 2k and 10k); thermocouples (types J, K and T) |
| Digital inputs | 4 digital inputs, configurable as: fast counters or frequency (30 kHz, 2 max), slow counters or frequency (100 Hz, 2 max), digital state (logic level / open collector / switch closure, 4 max) |
| Serial input channel | 62 SDI-12 sensors or a single WET Sensor |
| SDI-12 | Compliant with SDI-12 Specification Version 1.3; limits: 62 SDI-12 sensors, 50 PR2/6 SDI-12 Profile Probes, 300 measurements in the logging program (depending on complexity), 300 m cable (tested with 6 × PR2 SDI-12 sensors); regulated +12 V, 0.5 A switched supply |
| Accuracy conditions | Accuracy given at 25 °C and over -20 to +60 °C; noise figures are rms values |
| Voltage ±23 mV | 0.022 % + 12 µV at 25 °C; 0.08 % + 27 µV at -20 to +60 °C; noise 2.0 µV (selectable adaptive auto-ranging / fixed range) |
| Voltage ±185 mV | 0.009 % + 22 µV at 25 °C; 0.07 % + 38 µV at -20 to +60 °C; noise 2.5 µV |
| Voltage -1.4 to +1.5 V / -0.17 to +2.7 V | 0.005 % + 115 µV at 25 °C; 0.04 % + 150 µV at -20 to +60 °C; noise 25 µV (single-ended voltage measurements are susceptible to additional offset errors) |
| Thermocouple 0 to 70 °C | 0.39 °C at 25 °C; 0.79 °C at -20 to +60 °C; noise 0.05 °C; cold junction measured at isothermal terminals, response < 0.1 °C / °C/hour |
| Bridge ±7.5 mV/V | 0.05 % + 15 µV/V at 25 °C; 0.09 % + 37 µV/V at -20 to +60 °C; noise 1.5 µV/V (mV per 1 V excitation) |
| Bridge ±62 mV/V | 0.04 % + 24 µV/V at 25 °C; 0.08 % + 48 µV/V at -20 to +60 °C; noise 2 µV/V |
| Potentiometer 0 to 1 (ratiometric) | 0.036 % + 0.00015 at 25 °C; 0.057 % + 0.00017 at -20 to +60 °C; noise 0.00002 |
| Resistance 3-wire 1 kΩ | 0.1 % + 0.4 Ω at 25 °C; 0.21 % + 0.4 Ω at -20 to +60 °C; noise 0.15 Ω |
| Resistance 3-wire 9 kΩ | 0.07 % + 0.9 Ω at 25 °C; 0.19 % + 1 Ω at -20 to +60 °C; noise 0.2 Ω |
| Resistance 3-wire 135 kΩ | 0.05 % + 6 Ω at 25 °C; 0.14 % + 7 Ω at -20 to +60 °C; noise 1.0 Ω |
| Resistance 2-wire 9 kΩ | 0.06 % + 12 Ω at 25 °C; 0.16 % + 18 Ω at -20 to +60 °C; noise 0.2 Ω |
| Resistance 2-wire 135 kΩ | 0.05 % + 17 Ω at 25 °C; 0.11 % + 23 Ω at -20 to +60 °C; noise 1.0 Ω |
| Thermistor 2k, -20 to +60 °C | 0.05 °C at 25 °C; 0.08 °C at -20 to +60 °C; noise < 0.01 °C (3-wire resistance measurement) |
| Thermistor 10k, -20 to +60 °C | 0.04 °C at 25 °C; 0.09 °C at -20 to +60 °C; noise < 0.01 °C (3-wire resistance measurement) |
| Long-term stability | ±0.02 % worst case over 1 year |
| Input impedance | 0.8 to 3.8 GΩ |
| Common mode range | +3 V to -2.5 V |
| Common mode rejection ratio | > 70 dB |
| Input protection | All input terminals protected to ±15 V DC or 24 V AC continuous, including battery reverse polarity |
| Sensor excitation | Calibrated 3 V reference, +5 V and +12 V stabilised, or 5 to 10.5 V (battery or external power), user selectable |
| Recording options | Individual readings, statistics, total, integral, wind direction, vector average, gust, wind roses, conditional recording; conversion to engineering units by look-up tables, polynomial or linear conversion |
| Recording rate | 1 second to > 24 hrs, independently programmable for each channel |
| Sampling rate | Typically > 16 channels per second |
| Calculated measurements | Unlimited user-configured virtual channels using algebraic and trigonometric functions |
| Reading storage | 4 MB FLASH memory storing 2.5 million readings (typical), exported as text file |
| Control outputs | 2 relay outputs, expandable to 6 with the Relay Expansion Module (GP2-RLY); latching SPST relays rated 1 A, 24 V AC 32 V DC |
| Communications | SDI-12, RS-232 serial, 115.2 kBaud; USB adaptor cable included; cellular modem options for remote access |
| Networking | Up to 7 GP2s on 100 m of network cabling (M12) |
| Software | DeltaLINK PC software (supplied free): program editor, Script Editor, Simulator, data download and chart/table display; SDI-12 from DeltaLINK 3.2; Penman-Monteith ETo from version 3.7 |
| Internal power | 6 × AA alkaline cells, typically sufficient for 300k readings |
| External power | 10 to 15 V DC, 2 A via screw terminals or network cabling |
| Sleep current | < 60 µA typical + 30 µA for each digital input held low |
| Wake current | < 10 mA + any current supplied to sensors |
| Operating temperature | -20 to +60 °C |
| Enclosure | Weatherproof case, IP65, with desiccant and humidity indicator |
| EMC conformity | Tested to comply with EN 50081-1 and EN-50082-1 (1992) harmonised emissions and immunity standards |
| Dimensions | 225 × 185 × 75 mm |
| Weight | 1.0 kg (base configuration) |