Out of Stock
Industrial & Scientific

Onset HOBO U20L-01 Water Level and Temperature Data Logger, 0 to 9 m (30 ft) Depth, 0 to 207 kPa Absolute Pressure, Non-Vented, IP68

Brand: Generic
Out of Stock

The Onset HOBO U20L-01 is a non-vented water level logger that records absolute pressure and temperature in wells, streams, lakes, wetlands and tidal areas. It covers 0 to 207 kPa, about 0 to 9 m (30 ft) of water at sea level, with a typical water level error of ±0.1% FS (1.0 cm).

Onset HOBO U20L-01 Water Level Data Logger

The HOBO U20L-01 Water Level Logger is used for monitoring changing water levels in a wide range of applications, including streams, lakes, wetlands, tidal areas and groundwater. It records absolute pressure and temperature, and HOBOware Pro software converts the pressure readings into water level readings using a reference water level, with compensation for temperature, fluid density and barometric pressure.

The logger features a ceramic pressure sensor, a durable polypropylene housing and a protective end cap for deployment in existing wells or stilling wells. Because it uses an absolute pressure sensor, there are no vent tubes or desiccants to maintain. The U20L-01 covers 0 to 207 kPa (0 to 30 psia), approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, and with this smaller range it is also a good barometric reference for other water level loggers in the same area.

HOBOware Pro (version 3.5 or higher) is required for logger operation, together with a coupler (COUPLER2-C) and either the USB Optic Base Station (BASE-U-4) or the HOBO Waterproof Shuttle (U-DTW-1, firmware version 3.2.0 or later) for configuration and data offload. Note: a calibration certificate is not offered for this logger; if a NIST-traceable calibration certificate is required, see the HOBO U20 series.

Key Features

  • Absolute pressure range 0 to 207 kPa (0 to 30 psia): approximately 0 to 9 m (0 to 30 ft) of water depth at sea level
  • Water level accuracy: typical error ±0.1% FS, 1.0 cm (0.03 ft) water; maximum error ±0.2% FS, 2.0 cm (0.06 ft) water
  • Temperature measurement from -20° to 50°C with ±0.44°C accuracy from 0° to 50°C
  • Non-vented design: no vent tubes or desiccants to maintain
  • Ceramic pressure sensor, individually calibrated and temperature compensated from 0° to 40°C
  • Polypropylene housing and stainless steel screws suitable for saltwater; IP68 environmental rating
  • 64K bytes memory, approximately 21,700 pressure and temperature samples
  • Logging intervals from 1 second to 18 hours, with up to 8 user-defined logging intervals and durations
  • 3.6 Volt lithium battery with a typical life of 5 years at a logging interval of 1 minute or greater
  • Optical USB interface: data is offloaded without breaking the integrity of the seals

Optional Accessories

  • Cable (CABLE-1-300 or CABLE-1-50) and Cable Crimp (CABLE-1-CRIMP)
  • Replacement Coupler (COUPLER2-C)

Product Specifications

ManufacturerOnset Computer Corporation (HOBO Data Loggers)
ModelU20L-01
Pressure Operation Range (Absolute)0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude
Factory Calibrated Range69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F)
Burst Pressure310 kPa (45 psia) or 18 m (60 ft) depth
Water Level AccuracyTypical error: ±0.1% FS, 1.0 cm (0.03 ft) water; maximum error: ±0.2% FS, 2.0 cm (0.06 ft) water (with accurate reference water level measurement, known water density, accurate Barometric Compensation Assistant data and a stable temperature environment)
Raw Pressure Accuracy±0.3% FS, 0.62 kPa (0.09 psi) maximum error (includes all sensor drift, temperature and hysteresis-induced errors)
Pressure Resolution<0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water
Pressure Response Time (90%)<1 second at a stable temperature; allow 20 minutes in water to achieve full temperature compensation of the pressure sensor
Pressure Sensor DriftLess than 0.5% FS (worst case) per year
Temperature Operation Range-20° to 50°C (-4° to 122°F)
Temperature Accuracy±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F)
Temperature Resolution0.10°C at 25°C (0.18°F at 77°F)
Temperature Response Time (90%)10 minutes in water (typical)
Temperature Stability (Drift)0.1°C (0.18°F) per year
Real-time Clock±1 minute per month, 0° to 50°C (32° to 122°F)
Battery2/3 AA, 3.6 Volt lithium, factory-replaceable
Battery Life (Typical Use)5 years with 1 minute or greater logging interval
Memory (Non-volatile)64K bytes memory (approx. 21,700 pressure and temperature samples)
Logging IntervalFixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours
Launch ModesImmediate start and delayed start
Offload ModesOffload while logging; stop and offload
Battery IndicationBattery voltage can be viewed in status screen and optionally logged in datafile; low battery indication in datafile
InterfaceOptical, via coupler (COUPLER2-C) with USB Optic Base Station (BASE-U-4) or HOBO Waterproof Shuttle (U-DTW-1)
SoftwareHOBOware Pro, version 3.5 or higher
Wetted MaterialsPolypropylene housing and lanyard; Viton and Buna-N O-rings; ceramic sensor in acetal end cap; stainless steel screws suitable for saltwater
Environmental RatingIP68
ComplianceCE
Dimensions3.18 cm (1.25 inches) diameter, 15.24 cm (6.0 inches) length; mounting hole 6.3 mm (0.25 inches) diameter
WeightApproximately 154 g (5.43 oz) in air; approximately 53.9 g (1.9 oz) in fresh water