Four-channel digital thermometer for Type K/J/T/E/R/S thermocouples and Pt 100 ohm probes, showing all four channels at once. It logs readings with date and time to an SD card that opens directly in Excel, and offers T1-T2 difference, data hold and max/min recall.
The Lutron TM-947SD is a four-channel digital thermometer with a built-in real-time datalogger. It accepts Type K, J, T, E, R and S thermocouple probes as well as Pt 100 ohm probes, and shows the four channels (T1, T2, T3, T4) on the LCD at the same time. It can also display the temperature difference T1-T2. The measuring range is -100 to 1300 °C with a Type K probe and -199.9 to 850.0 °C with a Pt 100 ohm probe, in °C or °F, with 0.1 degree or 1 degree resolution depending on the range.
Thanks to its built-in clock and calendar, the meter saves every reading together with the date and time on an SD memory card (1 GB to 16 GB). The sampling time can be set from 1 second to 3600 seconds, and a manual mode saves one reading at each press of the logger button with a position (location) number from 1 to 99. No extra software is needed: the SD card is simply plugged into the computer and the recorded values open directly in Excel for further data or graphic analysis.
An RS232/USB computer interface is available for transferring live readings to a PC using the optional USB cable USB-01 or RS232 cable UPCB-02 and the optional Data Acquisition software SW-U801-WIN. Temperature probes are optional accessories and are ordered separately.
| Manufacturer | Lutron Electronic |
|---|---|
| Model | TM-947SD |
| Channels | T1, T2, T3, T4, T1-T2 |
| Sensor Type | Type K thermocouple probe; Type J/T/E/R/S thermocouple probe; Pt 100 ohm probe (0.00385 alpha coefficient, meets DIN IEC 751) |
| Resolution | 0.1 °C / 1 °C, 0.1 °F / 1 °F |
| Type K - Range, Resolution, Accuracy | -100.0 to -50.1 °C: 0.1 °C, ± (0.4 % + 1 °C); -50.0 to 999.9 °C: 0.1 °C, ± (0.4 % + 0.5 °C); 1000 to 1300 °C: 1 °C, ± (0.4 % + 1 °C). -148.0 to -58.1 °F: 0.1 °F, ± (0.4 % + 1.8 °F); -58.0 to 999.9 °F: 0.1 °F, ± (0.4 % + 1 °F); 1000 to 2372 °F: 1 °F, ± (0.4 % + 2 °F) |
| Type J - Range, Resolution, Accuracy | -100.0 to -50.1 °C: 0.1 °C, ± (0.4 % + 1 °C); -50.0 to 999.9 °C: 0.1 °C, ± (0.4 % + 0.5 °C). -148.0 to -58.1 °F: 0.1 °F, ± (0.4 % + 1.8 °F); -58.0 to 999.9 °F: 0.1 °F, ± (0.4 % + 1 °F) |
| Type T - Range, Resolution, Accuracy | -100.0 to -50.1 °C: 0.1 °C, ± (0.4 % + 1 °C); -50.0 to 400.0 °C: 0.1 °C, ± (0.4 % + 0.5 °C). -148.0 to -58.1 °F: 0.1 °F, ± (0.4 % + 1.8 °F); -58.0 to 752.0 °F: 0.1 °F, ± (0.4 % + 1 °F) |
| Type E - Range, Resolution, Accuracy | -100.0 to -50.1 °C: 0.1 °C, ± (0.4 % + 1 °C); -50.0 to 900.0 °C: 0.1 °C, ± (0.4 % + 0.5 °C). -148.0 to -58.1 °F: 0.1 °F, ± (0.4 % + 1.8 °F); -58.0 to 999.9 °F: 0.1 °F, ± (0.4 % + 1 °F); 1000 to 1652 °F: 1 °F, ± (0.4 % + 2 °F) |
| Type R - Range, Resolution, Accuracy | 0 to 600 °C and 601 to 1700 °C: 1 °C, ± (0.5 % + 1 °C). 32 to 1112 °F and 1113 to 3092 °F: 1 °F, ± (0.5 % + 2 °F) |
| Type S - Range, Resolution, Accuracy | 0 to 600 °C and 601 to 1500 °C: 1 °C, ± (0.5 % + 1 °C). 32 to 1112 °F and 1113 to 2732 °F: 1 °F, ± (0.5 % + 2 °F) |
| Pt 100 ohm - Range, Resolution, Accuracy | -199.9 to 850.0 °C: 0.1 °C, ± (0.4 % + 1 °C). -327.0 to 999.9 °F: 0.1 °F, ± (0.4 % + 1.8 °F); 1000 to 1562 °F: 1 °F, ± (0.4 % + 2 °F) |
| Accuracy Conditions | Specified for the meter only, at an environment temperature of 23 ± 5 °C |
| Display | LCD, 52 mm x 38 mm, with green backlight (ON/OFF) |
| Sampling Time of Display | Approx. 1 second |
| Datalogger Sampling Time | Auto: 1 second to 3600 seconds (at 1 second, memory data may be lost). Manual: one reading saved per press of the logger button (sampling time set to 0 second), with position no. 1 to 99 |
| Memory Card | SD memory card, 1 GB to 16 GB |
| Data Output | RS 232/USB PC computer interface (optional RS232 cable UPCB-02 or optional USB cable USB-01) |
| Probe Input Socket | Type K/J/T/E/R/S: 2 pin thermocouple socket; Pt 100 ohm: ear phone socket |
| Temperature Compensation | Automatic temperature compensation for Type K/J/T/E/R/S; linear compensation for the full range |
| Offset Adjustment | Available for Type K/J/T/E/R/S and Pt 100 ohm |
| Data Hold / Memory Recall | Freezes the display reading; maximum and minimum value recall |
| Power Off | Auto shut off or manual off by push button |
| Operating Temperature | 0 to 50 °C |
| Operating Humidity | Less than 85% R.H. |
| Power Supply | Alkaline or heavy duty DC 1.5 V battery (UM3, AA) x 6 PCs, or DC 9V adapter input (AC/DC power adapter is optional) |
| Power Current | Normal operation (no SD card saving, backlight OFF): approx. DC 8.5 mA; while saving data to the SD card (backlight OFF): approx. DC 30 mA; backlight ON adds approx. 14 mA |
| Standards | ISO-9001, CE, IEC1010 |
| Dimensions | 177 x 68 x 45 mm (7.0 x 2.7 x 1.9 inch) |
| Weight | 489 g / 1.08 LB |