Industrial & Scientific

Lutron YK-2001PHA Portable Multi-Parameter Water Meter: pH 0-14, ORP ±1999 mV, Conductivity to 200 mS, TDS, Salt 0-12 %, DO 0-20.0 mg/L

Brand: Lutron
14,400 EGP

Portable multi-purpose meter for pH/ORP, conductivity/TDS, dissolved oxygen and salt in one unit. Supplied with the CDPB-03 conductivity/TDS/salt probe; pH, ORP, dissolved oxygen and ATC probes are optional. Backlit LCD, data hold, max/min record and RS232/USB output.

Quantity:
– +

Lutron YK-2001PHA pH/ORP, DO, Conductivity/TDS and Salt Meter

The Lutron YK-2001PHA is one meter for multi-purpose operation: it measures pH/ORP, conductivity/TDS, dissolved oxygen and salt, each with its own separate probe so the instrument is easy to use in different measurement environments. It covers pH 0 to 14, ORP -1999 to 1999 mV, conductivity in four ranges (200 uS, 2 mS, 20 mS, 200 mS), salt from 0 to 12 % (% weight) and dissolved oxygen from 0 to 20.0 mg/L.

The meter is supplied with the CDPB-03 carbon rod conductivity/TDS/salt probe. The pH electrode (any pH electrode with BNC connector), ORP electrode, polarographic dissolved oxygen probe OXPB-11 and ATC temperature probe TP-07 are optional accessories. pH measurement supports automatic (with the ATC probe) or manual temperature compensation and auto calibration at pH 7, pH 4 and pH 10 or other values; conductivity measurement has automatic temperature compensation with a selectable temperature coefficient.

A microcomputer circuit, an LCD with green backlight, data hold, maximum/minimum recording and an RS232/USB PC interface make it suitable for water conditioning, aquariums, fish hatcheries, beverage and food processing, laboratories, the paper and plating industries, quality control and education.

Key Features

  • One meter for pH/ORP, conductivity/TDS, dissolved oxygen and salt measurement.
  • pH: 0 to 14 pH with 0.01 pH resolution; ORP: -1999 to 1999 mV.
  • Conductivity: 200 uS / 2 mS / 20 mS / 200 mS ranges, selectable uS/mS or TDS.
  • Dissolved oxygen: 0 to 20.0 mg/L with the optional polarographic oxygen probe with temperature sensor; the heavy duty probe head can connect with a BOD bottle.
  • Salt: 0 to 12 % salt (% weight).
  • pH function selects pH or ORP; ATC or manual temperature adjustment; auto calibration for pH 7, pH 4 and pH 10 or other value.
  • Automatic temperature compensation for conductivity with selectable temperature coefficient of the measured solution.
  • DO measurement with automatic temperature compensation and built-in "% SALT" and "Mountain Height" compensation value adjustment.
  • Separate probes, easy to operate in different measurement environments.
  • LCD with green backlight for easy reading.
  • Data hold, record of maximum and minimum readings.
  • Default auto power off or manual power off.
  • RS232/USB PC computer interface.
  • Powered by a DC 9V battery (006P) or a DC 9V adapter.

Package Contents

  • YK-2001PHA meter
  • Conductivity/TDS/Salt probe CDPB-03 — 1 PC
  • Instruction manual — 1 PC

Optional Accessories

  • Hard carrying case CA-06
  • pH electrodes PE-03, PE-11, PE-01, PE06HD, PE-04HD, PE-03K7
  • ATC (automatic temperature) probe TP-07
  • pH 7 buffer solution pH-07 and pH 4 buffer solution pH-04
  • ORP electrode ORP-14
  • Pt type conductivity/TDS/salt probe CDPB-04
  • 1.413 mS conductivity standard solution CD-14
  • Oxygen probe OXPB-11, spare probe head with diaphragm set OXHD-04, probe-filling electrolyte OXEL-03
  • AC to DC 9V adapter
  • USB cable USB-01, RS232 cable UPCB-02
  • Data acquisition software SW-U801-WIN, Excel data acquisition software SW-E802

Product Specifications

ManufacturerLutron Electronic
ModelYK-2001PHA
Measurement FunctionspH/ORP, Conductivity/TDS (Total Dissolved Solids), Dissolved Oxygen, Salt
pH Range0 to 14 pH
pH Resolution0.01 pH
pH Accuracy± (0.02 pH + 2 d) — based on calibrated meter only
ORP (mV) Range-1999 to 1999 mV
ORP Resolution1 mV
ORP Accuracy± (0.5% + 2 d)
pH ElectrodeOptional — any pH electrode with BNC connector
pH Input Impedance10^12 ohm
pH Temperature CompensationManual: 0 to 100 °C, adjusted by push button on the front panel; Automatic: 0 to 65 °C with the optional ATC probe TP-07
pH CalibrationpH 7, pH 4 and pH 10 — 3 points calibration
Conductivity Ranges200 uS: 0 to 200.0 uS; 2 mS: 0.2 to 2.000 mS; 20 mS: 2 to 20.00 mS; 200 mS: 20 to 200.0 mS
Conductivity Resolution0.1 uS / 0.001 mS / 0.01 mS / 0.1 mS (by range)
Conductivity Accuracy± (2 % F.S. + 1 d), specified for measurement values ≤ 100 mS, @ 23 ± 5 °C
TDS Ranges200 PPM: 0 to 132 PPM; 2,000 PPM: 132 to 1,320 PPM; 20,000 PPM: 1,320 to 13,200 PPM; 200,000 PPM: 13,200 to 132,000 PPM
TDS Resolution0.1 PPM / 1 PPM / 10 PPM / 100 PPM (by range)
TDS Accuracy± (2 % F.S. + 1 d), specified for measurement values ≤ 66,000 PPM, @ 23 ± 5 °C
Temperature (Conductivity Probe)0 to 60 °C (32 to 140 °F); resolution 0.1 °C / 0.1 °F; accuracy ± 0.8 °C / ± 1.5 °F
Salt Range0 to 12 % salt (% weight)
Salt Resolution0.01 % salt
Salt Accuracy± 0.5 % salt value
Temperature Compensation (Conductivity/TDS/Salt)Automatic from 0 to 60 °C (32 - 140 °F), compensation factor variable between 0 and 5.0 % per °C
Conductivity/TDS/Salt ProbeCDPB-03 (included), carbon rod electrode; round, 22 mm dia. x 120 mm length; probe operating temperature 0 to 60 °C
Dissolved Oxygen ProbeOptional OXPB-11, polarographic type; probe size 190 mm x 28 mm dia.
Dissolved Oxygen0 to 20.0 mg/L; resolution 0.1 mg/L; accuracy ± 0.4 mg/L
Oxygen in Air0 to 100.0 %; resolution 0.1 % O2; accuracy ± 0.7 % O2
Temperature (DO Probe)0 to 50 °C; resolution 0.1 °C; accuracy ± 0.8 °C / 1.5 °F
DO Compensation & AdjustmentTemperature: 0 to 50 °C, automatic; Salt: 0 to 50 % salt; Height: 0 to 8900 meter
DisplayLCD, 52 mm x 38 mm, with green backlight (ON/OFF)
Sampling Time of DisplayApprox. 1 second
Data Hold / MemoryData hold; recall of maximum and minimum values
Data OutputRS232/USB PC computer interface (with the optional RS232 cable UPCB-02 or USB cable USB-01)
Operating Temperature0 to 50 °C
Operating HumidityLess than 85% R.H.
Power SupplyAlkaline or heavy duty DC 9V battery (006P) or equivalent; DC 9V adapter input (AC/DC adapter is optional)
Power CurrentApprox. DC 14 mA in normal operation (backlight OFF); the backlight adds approx. 12 mA
StandardsISO-9001, CE, IEC1010
Dimensions177 x 68 x 45 mm (7.0 x 2.7 x 1.9 inch)
Weight489 g / 1.08 LB