The Elmetron CC-411 is a portable, IP-66 waterproof meter for conductivity, salinity, TDS and temperature. It measures up to 199.9 mS/cm over four auto-switched sub-ranges, shows the function and temperature together, stores 50 results and runs on a 9 V battery.
The Elmetron CC-411 is a simple portable meter that measures conductivity, salinity, TDS and temperature, for use in the field and in the laboratory. It measures liquids with conductivity up to 200 mS/cm over 4 sub-ranges switched automatically, and its display shows the measured function and the temperature at the same time.
Salinity is calculated from conductivity according to the real characteristic rather than a constant coefficient, with conversion to NaCl or KCl, and the total dissolved solids (TDS) value is estimated with a coefficient set by the user between 0.20 and 1.00. The cell constant K can be entered directly or determined with a standard solution. Temperature compensation is automatic with a constant α coefficient of 2 %/°C and a reference temperature of 25 °C or 20 °C (e.g. for measurements in honey); for precise temperature-compensated measurements Elmetron recommends its 401 or 461 series meters.
Low weight, small dimensions and a waterproof IP-66 housing make the CC-411 well suited to work in difficult field conditions. With a special sensor it can also measure the electric admittance of tree seedlings to determine their vitality. The meter has no PC connection.
| Manufacturer | Elmetron |
|---|---|
| Model | CC-411 |
| Measured parameters | Conductivity, salinity, TDS, temperature |
| Conductivity range | 0 to 199.9 mS/cm, auto-range (4 sub-ranges) |
| Conductivity accuracy (± 1 digit) | ±0.25 % up to 19.99 mS/cm; ±0.5 % above 20 mS/cm (meter only) |
| Cell constant K | 0.010 to 1.999 cm⁻¹ |
| Salinity | Conversion to NaCl or KCl; KCl up to 140 g/l |
| Salinity accuracy (± 1 digit) | ±2 % (meter only) |
| TDS | Estimated value, coefficient 0.20 to 1.00 set by user |
| Temperature range | -50.0 to 199.9 °C |
| Temperature accuracy (± 1 digit) | ±0.1 °C (meter only); acceptable probe error in 0 to 100 °C: ±0.8 °C with Pt-1000B, ±0.35 °C with Pt-1000A |
| Temperature compensation | Automatic, -5 to 70 °C |
| α coefficient | Constant 2 %/°C |
| Reference temperature | 25 °C or 20 °C |
| Temperature sensor | Pt-1000B |
| Display | Simultaneous readout of measured function and temperature |
| Memory | 50 results |
| Housing | Waterproof, IP-66 |
| Power supply | 9 V battery; 12 V power adapter as option |
| Auto power-off | Yes |
| Dimensions (L × W × H) | 149 × 82 × 22 mm |
| Weight | 220 g |