Benetech GM3125 digital insulation tester with 500, 1000, 2500 and 5000 V test voltages and a range up to 1000 GΩ. It also measures AC/DC voltage up to 600 V and offers PI and timer tests, auto-discharge, live-circuit warning and a backlit LCD display.
The Benetech GM3125 is a digital insulation resistance tester with four selectable DC test voltages – 500 V, 1000 V, 2500 V and 5000 V – measuring insulation resistance up to 1000 GΩ on the 5000 V range. Insulation resistance accuracy is ±5% rdg ±3 dgt from 0 to 99.9 GΩ and ±20% rdg ±3 dgt above 100 GΩ.
Besides insulation testing, the instrument measures DC voltage from ±30 to ±600 V and AC voltage from 30 to 600 V (50/60 Hz) with 1 V resolution and ±2% rdg ±3 dgt accuracy, and provides polarization index (PI) measurement and a timer measurement function.
Readings appear on a backlit LCD of max. 999 counts with a bar graph of max. 36 points. An auto-discharge function and live-circuit warning symbols with an audible alarm make operation safer. The tester runs on 8 × 1.5 V LR14 batteries (approx. 15 hours) or on a 12 V DC adapter (12 V/1 A).
| Manufacturer | Benetech |
|---|---|
| Model | GM3125 |
| Insulation test voltages (rated) | 500 V / 1000 V / 2500 V / 5000 V |
| Insulation resistance range at 500 V | 0.0–99.9 MΩ; 100–999 MΩ |
| Insulation resistance range at 1000 V | 0.0–99.9 MΩ; 100–999 MΩ; 1.00–1.99 GΩ |
| Insulation resistance range at 2500 V | 0.0–99.9 MΩ; 100–999 MΩ; 1.00–9.99 GΩ; 10.0–99.9 GΩ |
| Insulation resistance range at 5000 V | 0.0–99.9 MΩ; 100–999 MΩ; 1.00–9.99 GΩ; 10.0–99.9 GΩ; 100–1000 GΩ |
| Open-circuit voltage | DC 500 V +30% −0%; DC 1000 V +20% −0%; DC 2500 V +20% −0%; DC 5000 V +20% −0% |
| Rated current | 1 mA – 1.2 mA |
| Short-circuit current | Approx. 1.3 mA |
| Insulation resistance accuracy | ±5% rdg ±3 dgt (0–99.9 GΩ); ±20% rdg ±3 dgt (above 100 GΩ) |
| DC voltage range | ±30 – ±600 V |
| AC voltage range | 30 – 600 V (50/60 Hz) |
| Voltage resolution | 1 V |
| Voltage accuracy (DC and AC) | ±2% rdg ±3 dgt |
| Display | Backlit LCD, max. 999 counts (1000 counts displayed only at 1 TΩ); bar graph max. 36 points |
| Over-range indication | OL on insulation resistance range; LO on voltage range |
| Auto-ranging | Shifts to upper range at 1000 counts, to lower range at 95 counts (insulation resistance range only) |
| Sample rate | 0.5 – 10 times/s |
| Overload protection | Insulation resistance: AC 1200 V / 10 s; Voltage: AC 720 V / 10 s |
| Insulation resistance (circuit to enclosure) | 1000 MΩ or more at DC 1000 V |
| Power supply | DC 12 V (8 × 1.5 V LR14 battery); suitable for 12 V DC adapter (12 V/1 A) |
| Battery life | Approx. 15 hours |
| Auto power off | After 10 minutes without operation |
| Operating altitude | Less than 2000 m (indoor use) |
| Operating conditions | 0 – 40 °C, humidity ≤ 85% |
| Storage conditions | -20 – 60 °C, humidity ≤ 90% |
| Dimensions | 153 × 96 × 220 mm |
| Weight | 1032 g (without batteries and test wires) |